Operating Manual 3011
Digital indicator 3011
Operating Manual Digital Indicator 3011
Version V0.1
Stand: 06.09.07
Page 1 of 8
Subject to technical modifications. Drafts are protected according to §823 BGB: Copy, duplication or circulation only with explicit approval.
Soehnle Professional GmbH & Co. KG
|
P.O. Box 1308
|
D-71536 Murrhardt | Telephone +49 (0) 7192 / 9319 – 0
|
E-Mail [email protected]
4.Y-8328
Operating Manual 3011
Electronic nameplate type 3011
By pressing the keys
and
, the electronic nameplate is indicated.
T
• With the arrow keys (
), the electronic nameplate as well as the
T F
serial number and calibration counter of the measuring point can be indicated on the
display.
The indication of the electronic nameplate is terminated by pressing the key
For a one-range scale, the parameters of the other range are not applicable.
.
Indication
Stepps
Parameters
Display
1.
2.
3.
4.
5.
6.
7.
8.
9.
Calibration counter
Type
Year
Serial number
Ident- Chip
Max of range 1
Min of range 1
e of range 1
Max of range 2
Min of range 2
e of range 2
E 2
T 2880
Y 07
1234
Id 1
¯ 10,00kg
_
0,20kg
0,01kg
¯ 20,00kg additionally sign for range 2
E
10.
11.
_
E
0,40kg additionally sign for range 2
0,02kg additionally sign for range 2
For non-approvable scales, “d“ instead of “E“ is indicated.
Operating Manual Digital Indicator 3011
Version V0.1
Stand: 06.09.07
Page 3 of 8
Subject to technical modifications. Drafts are protected according to §823 BGB: Copy, duplication or circulation only with explicit approval.
Soehnle Professional GmbH & Co. KG
|
P.O. Box 1308
|
D-71536 Murrhardt | Telephone +49 (0) 7192 / 9319 – 0
|
E-Mail [email protected]
4.Y-8328
Operating Manual 3011
SWITCH On/Off Type 3011
• Switch the indicator on/off via the ON/OFF key.
• While switching on, the scale is zeroed. If the scale is switched on outside the zero range
at power-on, the scale indicates an error message:
_0_ = Measuring value below zero limit
_ _
0 = Measuring value above zero limit
The application is switched off to avoid the cause for under/overload before the next
switch on. The zero range at power-on can be skipped by pressing the zero key during
the switch on routine. The last saved zero point is accepted.
The active measuring point is indicated as flag on the display. Switchover to another
measuring point is realised by pressing the Info and Print key on the keyboard.
• Zero range limits at power-on:
Approvable: Zero range at power-on 20% of maximum, default –5% to +15% of maximum
Non-approvable: Zero range at power-on –99% to +99% of maximum
• Behaviour at power failure: the scale returns in the last saved state.
• At power-on (warm or cold starting), the last used program is activated.
Operating Manual Digital Indicator 3011
Version V0.1
Stand: 06.09.07
Page 4 of 8
Subject to technical modifications. Drafts are protected according to §823 BGB: Copy, duplication or circulation only with explicit approval.
Soehnle Professional GmbH & Co. KG
|
P.O. Box 1308
|
D-71536 Murrhardt | Telephone +49 (0) 7192 / 9319 – 0
|
E-Mail [email protected]
4.Y-8328
Operating Manual 3011
Zeroing Type 3011
• Zeroing via the zero key.
• Zeroing approvable: -1 to +3% according to Approval Directive.
• Zero range non-approvable: Same range as set zero range limits at power-on.
• The display indicates after keypress
• at <-1% message Err 5
• at > 3% message Err 5
• Zero tracking: in the setting mode, the zero tracking can be switched on or off (on = 0,5
d/sec at standstill).
Operating Manual Digital Indicator 3011
Version V0.1
Stand: 06.09.07
Page 5 of 8
Subject to technical modifications. Drafts are protected according to §823 BGB: Copy, duplication or circulation only with explicit approval.
Soehnle Professional GmbH & Co. KG
|
P.O. Box 1308
|
D-71536 Murrhardt | Telephone +49 (0) 7192 / 9319 – 0
|
E-Mail [email protected]
4.Y-8328
Operating Manual 3011
Tara functions Type 3011
• The following tare function is available as standard fort he indicator 3011 to save
weighing data in the tare memory:
1. Manual tare
• Keypad management:
• Manual tare: Via the key "Tara"
on the membrane keyboard.
T
• Manual tare: by pressing the tare key. The tare value is saved, the weight indication is at
zero, switchover from gross to net is realised => “net” in display via a red LED. If the scale
is discharged, the tare weight shows a negative sign. Pressing the tare key again
overwrites the existing value, which means that a newly placed tare weight on the scale is
saved or the discharged scale with a negative tare weight on the display is zeroed.
• Delete tare memory:
• Pres the zero key.
Operating Manual Digital Indicator 3011
Version V0.1
Stand: 06.09.07
Page 6 of 8
Subject to technical modifications. Drafts are protected according to §823 BGB: Copy, duplication or circulation only with explicit approval.
Soehnle Professional GmbH & Co. KG
|
P.O. Box 1308
|
D-71536 Murrhardt | Telephone +49 (0) 7192 / 9319 – 0
|
E-Mail [email protected]
4.Y-8328
Operating Manual 3011
X10 DISPLAY Type 3011
• The tenfold higher resolution on the display is activated by pressing the keys “PRINT”
and “-0-“.
• If the scale is calibrated approvable, the tenfold higher resolution is indicated after a
keypress for 5 seconds or as long as the keys are hold.
Operating Manual Digital Indicator 3011
Version V0.1
Stand: 06.09.07
Page 7 of 8
Subject to technical modifications. Drafts are protected according to §823 BGB: Copy, duplication or circulation only with explicit approval.
Soehnle Professional GmbH & Co. KG
|
P.O. Box 1308
|
D-71536 Murrhardt | Telephone +49 (0) 7192 / 9319 – 0
|
E-Mail [email protected]
4.Y-8328
Operating Manual 3011
PRINT Type 3011
• An optional interface RS232 must be assembled in the digital indicator 3011 for the
function „Print“.
• The stored print template is send via the interface by pressing the print key.
Print template design and print conditions are realised via the service program.
Operating Manual Digital Indicator 3011
Version V0.1
Stand: 06.09.07
Page 8 of 8
Subject to technical modifications. Drafts are protected according to §823 BGB: Copy, duplication or circulation only with explicit approval.
Soehnle Professional GmbH & Co. KG
|
P.O. Box 1308
|
D-71536 Murrhardt | Telephone +49 (0) 7192 / 9319 – 0
|
E-Mail [email protected]
4.Y-8328
|